| Catalog Number |
0375941000 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the sheath broke off.No pieces fell into the joint space and the procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Alleged failure: metal became detached from the hood of the burr sheath the failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be the excessive force applied to the bur head, causing the flutes to rub on the housing surface.Also, a possibility of the bur tip accidentally encountered hard object during use.Based on the amount of damage observed, it is likely that excessive pressure was maintained for a substantial amount of time and poor suction.
|
| |
|
Event or Problem Description
|
|
It was reported that the sheath broke off.No pieces fell into the joint space and the procedure was completed successfully.
|
| |
|