• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG., BARREL BUR, 6 FLUTE F - SERIES - 4.0MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG., BARREL BUR, 6 FLUTE F - SERIES - 4.0MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 0375941000
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the sheath broke off.No pieces fell into the joint space and the procedure was completed successfully.
 
Additional Manufacturer Narrative
The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Alleged failure: metal became detached from the hood of the burr sheath the failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be the excessive force applied to the bur head, causing the flutes to rub on the housing surface.Also, a possibility of the bur tip accidentally encountered hard object during use.Based on the amount of damage observed, it is likely that excessive pressure was maintained for a substantial amount of time and poor suction.
 
Event or Problem Description
It was reported that the sheath broke off.No pieces fell into the joint space and the procedure was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG., BARREL BUR, 6 FLUTE F - SERIES - 4.0MM
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key7949809
Report Number0002936485-2018-00637
Device Sequence Number1381279
Product Code GFA
UDI-Device Identifier37613327054744
UDI-Public37613327054744
Combination Product (Y/N)N
Initial Reporter CountryGB
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0375941000
Device Lot Number18178CE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2018
Initial Date Received by Manufacturer 09/17/2018
Supplement Date Received by Manufacturer09/17/2018
Initial Report FDA Received Date10/09/2018
Supplement Report FDA Received Date12/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-