| Catalog Number |
10120 |
| Health Effect - Clinical Code |
Death (1802)
|
| Date of Event |
09/13/2018
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Investigation: the rdf was analyzed for this event.Review of the rdf does not show any indications of any issues during the procedure that would have contributed to the patient's death.Signals from the dlog showed no unusual alarms or abnormal procedure behavior.Given this information, there were no signals in the dlog that would indicate the machine did not operate as intended.The system was operating within the safety limits programmed in the machine.The device was found to be operating as intended, and within the generally tolerated limits for the majority of patients.There was no system or device malfunction identified that would indicate an issue with the optia system.The optia system appeared to be operating as intended and is safe to use.Investigation is in process, a follow-up report will be provided.
|
| |
|
Event or Problem Description
|
|
The customer reported that post a mono nuclear cell (mnc) collection on the spectra optia, the patient expired.Per the customer, they are unsure if the death was related to the procedure.The patient's vascular access was a central line.Patient age is not available at this time.The patient's gender was obtained from the run data file (rdf).The lot number, manufacture date and expiry date are not available at this time.The spectra optia collection set is not available for return because it was discarded by the customer.This report is being filed due to patient death, although per current information there is no suspected malfunction with the terumo bct device or allegation of a malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: a review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.
|
| |
|
Event or Problem Description
|
|
Per customer, no autopsy was not performed on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide corrected udi and additional information and investigation: the disposable set was not available for return.A full system checkout was performed on nov 13, 2018.The inspection confirmed that there was no device malfunction, and the device was released for use.Root cause: review of the d log does not show any indications of any issues during the procedure that would have contributed to the patient's death.Signals from the d log showed no unusual alarms or abnormal procedure behavior.Given this information, there were no signals in the d log that would indicate the machine did not operate as intended.The system was operating within the safety limits programmed in the machine.The device was found to be operating as intended, and within the generally tolerated limits for the majority of patients.There was no system or device malfunction identified that would indicate an issue with the optia system.The optia system appeared to be operating as intended and is safe to use.The physician treated the patient confirmed that the cause of death was due to lung embolism related to the patient's disease state.Per terumo bct's medical review, the device did not cause or contribute to this incident.
|
| |
|