• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. BENCHMARK 6F 071 DELIVERY CATHETER; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. BENCHMARK 6F 071 DELIVERY CATHETER; DQY Back to Search Results
Catalog Number BMK6F115M
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the benchmark was ovalized approximately 114.0 thru 121.0 cm from the hub.Conclusions: evaluation of the returned benchmark revealed that the distal tip was ovalized.If the peel-able sheath (supplied by penumbra) is not used during insertion of the benchmark into a parent catheter, damage such as ovalization may occur.The ovalization in the benchmark likely contributed to resistance experienced during insertion of the benchmark into the neuron max.The neuron max identified in the complaint was not returned to penumbra for evaluation.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a medical procedure using a benchmark 6f 071 delivery catheter (benchmark).During the procedure, while attempting to insert a benchmark into a neuron max 6f 088 long sheath (neuron max), the physician experienced resistance and just the tip of the benchmark would only advance into the neuron max; therefore, the benchmark was removed.The procedure was completed using another benchmark and the same neuron max.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENCHMARK 6F 071 DELIVERY CATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7949933
Report Number3005168196-2018-01986
Device Sequence Number857357
Product Code DQY
UDI-Device Identifier00814548014104
UDI-Public00814548014104
Combination Product (Y/N)Y
Initial Reporter CountryUK
PMA/510(K) Number
K142321
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/10/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/03/2021
Device Catalogue NumberBMK6F115M
Device Lot NumberF81274
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2018
Initial Report FDA Received Date10/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-