Catalog Number VA8078
Medical Device Problem Codes Retraction Problem (1536); Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.
 
Event or Problem Description
It was reported that during an angioplasty procedure in an av graft, the pta balloon allegedly ruptured, and sequentially had issues retracting through the sheath.The balloon and sheath were removed together as one unit.Another device was used to complete the procedure.There was no reported patient injury.
 
Event or Problem Description
It was reported that during an angioplasty procedure in an av graft, the pta balloon allegedly ruptured, and sequentially had issues retracting through the sheath.The balloon and sheath were removed together as one unit.Another device was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.The device was returned for evaluation.An inflation attempt was made and water was seen exiting the device from a longitudinal balloon rupture.Therefore, the investigation is confirmed for a longitudinal rupture.However, the investigation is inconclusive for the reported retraction issue as there were no signs of retraction issues on the returned device, and as the device was unable to be functionally tested due to the rupture.It is possible that a rupture could have resulted in the retraction issues through the sheath.However, the definitive root cause for the identified rupture or reported retraction issues could not be determined based on the available information.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(expiry date 7/2021).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VACCESS PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7950148
Report Number2020394-2018-01888
Device Sequence Number13645755
Product Code DQY
UDI-Device Identifier00801741059933
UDI-Public(01)00801741059933
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K111850
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVA8078
Device Lot Number93UC0052
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2018
Initial Date Received by Manufacturer 09/17/2018
Supplement Date Received by Manufacturer02/14/2019
Initial Report FDA Received Date10/09/2018
Supplement Report FDA Received Date03/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age55 YR
-
-