• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION ASCENDA; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION ASCENDA; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8785
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare professional (hcp) via a manufacturer representative (rep) regarding a patient who was receiving 1 mg/ml morphine at 48 mcg/day via an implantable pump for an unknown indication for use.It was reported that one collet fell off the end of the connector piece when it was taken from the kit.No environmental, external, or patient factors were noted to have led or contributed to the issue.No diagnostics or troubleshooting was performed.The collet was replaced with a connector from a new kit and the issue was resolved.The patient's status was "alive - no injury".No further issues were reported or anticipated.
 
Additional Manufacturer Narrative
Analysis identified the collet was missing from the catheter to catheter connector.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASCENDA
Common Device Name
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
MDR Report Key7950355
Report Number3007566237-2018-02982
Device Sequence Number1526132
Product Code LKK
UDI-Device Identifier00643169561601
UDI-Public00643169561601
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P860004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/20/2020
Device Model Number8785
Device Catalogue Number8785
Device Lot NumberN789247
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2018
Supplement Date Received by Manufacturer10/29/2018
Initial Report FDA Received Date10/09/2018
Supplement Report FDA Received Date11/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
-
-