| Model Number |
MMT-722NAS |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/07/2008
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Unit alarmed unexpected restart after battery change due to broken c37 solder joint on i/b.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump had a pump error unexpected restart.A cold reset was performed.The blood glucose at the time of the incident was 200 mg/dl.The device will be returned for analysis.
|
| |
|