Model Number 3.0NEO
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code Extubate (2402)
Date of Event 09/10/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the device had probably a thick and rigid material that caused a very painful insertion.It was also stated that the device was applied with bigger pressure, which also the reason of the problem.The customer indicated that the patient was re-intubated.
 
Additional Manufacturer Narrative
Evaluation summary: one sample was received for evaluation.No reported event was observed since met with the product specifications.A visual inspection was performed, and it was observed all the components are assembled properly at the tube according to drawing, a dimensional test was performed outer and inner diameter of the cannula was measured, this reading is within specification.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the change of tracheostomy tube, the insertion of the cannula was very painful due to the thick and rigid material.The device needed more pressure to be applied which may also be the reason of the problem.The patient was re-cannulated.Additional information indicates that after a third attempt to introduce the 3rd cannula, a cannula of a different lot number was used without issue.In subsequent insertion of 2 cannulas of the reported lot number, the painfulness and suffocation status were repeated.Also, by checking and by touch control she could feel the difference of the material.The patient has had a permanent tracheostomy.For over 2 years, the usage of the tracheostomy tubes is on a regular basis and requires very often changes of tubes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV(USD)
ave henequen no 1181 desarrol
ciudad juarez,mx 32590
MX  32590
MDR Report Key7951434
Report Number2936999-2018-00626
Device Sequence Number203329
Product Code JOH
UDI-Device Identifier30884522026653
UDI-Public30884522026653
Combination Product (Y/N)N
Initial Reporter CountryLO
PMA/510(K) Number
K945513
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/23/2020
Device Model Number3.0NEO
Device Catalogue Number3.0NEO
Device Lot Number15B0657JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2019
Initial Date Received by Manufacturer 09/12/2018
Supplement Date Received by Manufacturer06/13/2019
Initial Report FDA Received Date10/10/2018
Supplement Report FDA Received Date06/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age4 YR
-
-