| Model Number |
15 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio-control observed that the device had logged an event code in its memory which is related to a potential failure of the device to deliver defibrillation energy.Physio-control then replaced the device's main keypad assembly to resolve the reported issue related to the logged event code; however the cause of the "abnormal energy delivered" message could not be determined.After observing proper device operation through functional and performance testing the unit was returned to the customer for use.Based on the information available, physio-control has determined that it¿s reasonable to conclude that the likely cause of the reported issue was that the shock switch located on the main keypad assembly was out of tolerance due to excessive resistance.When this issue occurs, an event code will be logged in the device's memory following a failure of the device to deliver defibrillation energy.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device displayed "abnormal energy delivered".As a result, the wrong defibrillation or no defibrillation therapy would be delivered, if it were necessary.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
This product complaint requires a supplemental mdr to document that field action number 301587611/18/2019-001c is relevant to the reported issue.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device displayed "abnormal energy delivered".As a result, the wrong defibrillation or no defibrillation therapy would be delivered, if it were necessary.There was no patient use associated with the reported event.
|
| |
|