| Model Number |
AR40E |
| Medical Device Problem Code |
Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable no patient involvement reported.If explanted, give date: not applicable no patient involvement reported.(b)(6).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that after opening the lens case, it was noticed that the haptic of the intraocular lens (iol) was detaching.No patient involvement was reported.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint could not be verified.Manufacturing record review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specifications.A search revealed that no similar complaints have been received for this production order.Labeling review: the dfu adequately provides instructions and precautions for the proper use and handling of the product.The dfu states to examine the lens thoroughly to ensure particles have not become attached to it, and examine the lens optical surfaces for other defects.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|
Search Alerts/Recalls
|