U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV(USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV(USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 10LPC
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Information (3190)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the device had cuff leakage.Patient outcome was asked but unknown.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Evaluation summary: one sample device was received for evaluation.The reported condition was confirmed.An inflation/deflation test was performed and it was observed the cuff deflated after few seconds.A visual inspection was performed and it was observed the cuff presents a small tear.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the device had cuff leakage.Patient outcome was asked but unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV(USD)
ave henequen no 1181 desarrol
ciudad juarez,mx 32590
MX  32590
MDR Report Key7952253
Report Number2936999-2018-00637
Device Sequence Number1311624
Product Code JOH
UDI-Device Identifier60884522007110
UDI-Public60884522007110
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K962173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Model Number10LPC
Device Catalogue Number10LPC
Device Lot Number17C0752JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2019
Initial Date Received by Manufacturer 09/14/2018
Supplement Date Received by Manufacturer06/21/2019
06/21/2019
12/11/2019
Initial Report FDA Received Date10/10/2018
Supplement Report FDA Received Date07/09/2019
07/10/2019
12/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-