• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO MEDICAL CORPORATION GLIDESHEATH SLENDER NITINOL; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO MEDICAL CORPORATION GLIDESHEATH SLENDER NITINOL; INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was not implanted.Device was not explanted.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.A review of the device history record of the product code/lot# combination was conducted with no findings.With no device return the exact cause of the reported event cannot be definitively determined based on the available information.
 
Event or Problem Description
The user facility reported that after taking out the dilator of the sheath of the involved glideseath slender, blood was pulsating back through the valve as if the valve was not there.The doctor kept his thumb over the valve so that the bleeding would stop.The doctor then put a wire in quickly to exchange for a different sheath.After manipulating the wire, the valve seemed to get in position and the leakage stopped.The estimated blood loss was less than 250cc.There was no harm to the patient.The procedure was finished.Additional information was received on october 1, 2018.The procedure being performed was a radial left heart catheterization.The doctor did not change the sheaths, he wiggled it, while he had the wire in, and it stopped bleeding (it was believed that the valve realigned itself).Once bleeding was down to trickle, he proceeded with the rest of case.The patient was stable and the procedure was finished successfully.Additional information was received on october 3, 2018.After manipulating the wire, the valve seemed to get into position, and stopped the leakage.The initial sheath was kept in.Wires, balloons , and catheters were used the rest of the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESHEATH SLENDER NITINOL
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
MDR Report Key7952451
Report Number1118880-2018-00146
Device Sequence Number16526831
Product Code DYB
UDI-Device Identifier00389701011981
UDI-Public00389701011981
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K142183
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Model NumberN/A
Device Catalogue Number60-1065
Device Lot NumberWG06
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2018
Initial Report FDA Received Date10/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-