Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The complaint device has been returned, but the device investigation has not yet been completed.Once the evaluation is completed, a supplemental medwatch will be submitted.
|
|
Event Description
|
It was reported that the sales representative found the device fractured due to unknown reason.There was no reported adverse event.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows the repeated use and cracked on lateral side of post.Dhr was reviewed and no discrepancies relevant to the reported event were found.The root cause of the failure is considered to be a wear and tear from use.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|