Catalog Number A2015-080
Medical Device Problem Codes Material Rupture (1546); Difficult to Advance (2920)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Visual and functional inspection was performed on the returned device.The reported balloon rupture was confirmed during analysis of the returned device.The resistance was unable to be confirmed as it was based on case circumstances.Based on the information provided, the reported difficulties appear to be due to operational circumstances of the procedure.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Review of the complaint history of the reported lot revealed no other incidents.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that some resistance was met with the calcified anatomy during advancement of the 1.5x80 mm armada 14 to the heavily calcified left popliteal artery.During the first inflation, the armada 14 ruptured before it had reached nominal pressure.The balloon rupture was thought to be related to the heavy calcium.The armada 14 was removed in one piece without incident.A 3.0 mm armada 14 was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.There was no additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 14 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7952930
Report Number2024168-2018-07825
Device Sequence Number1386169
Product Code LIT
UDI-Device Identifier08717648155925
UDI-Public08717648155925
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K102705
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 10/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Catalogue NumberA2015-080
Device Lot Number6021241
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2018
Initial Report FDA Received Date10/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age73 YR
Patient Weight105
-
-