| Model Number |
TRUETRACK |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer on 10/1/2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 214, 216, 275, 286 and 263 mg/dl.The customer's expected fasting blood glucose test result range is 125 - 140 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was performed by the customer fasting and produced test results of 215 mg/dl and 205 mg/dl using the truetrack meter.Customer stated the product is stored in her bedroom and that she has no a/c in her home and uses a fan and opens the window in her room for cooling.The test strip lot manufacturer's expiration date is 08/21/2020 and open vial date is three weeks.The meter memory was reviewed for previous test result history (date/time not set as customer stated she changed the battery): (b)(6).
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 214, 216, 275, 286 and 263 mg/dl.The customer's expected fasting blood glucose test result range is 125 - 140 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was performed by the customer fasting and produced test results of 215 mg/dl and 205 mg/dl using the truetrack meter.Customer stated the product is stored in her bedroom and that she has no a/c in her home and uses a fan and opens the window in her room for cooling.The test strip lot manufacturer's expiration date is 08/21/2020 and open vial date is three weeks.The meter memory was reviewed for previous test result history (date/time not set as customer stated she changed the battery): result 1 :214mg/dl date:(b)(6) 2018, time:11:48am fasting, result 2 :216mg/dl date:(b)(6) 2018, time:11:58am fasting, result 3 :275mg/dl date:(b)(6) 2018, time:12:04pm fasting, result 4 :286mg/dl date:(b)(6) 2018, time:9:45am fasting, result 5 :263mg/dl date:(b)(6) 2018, time:9:45am fasting.
|
| |
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) internal report (b)(4).Returned meter and returned test strips evaluated with no defects found.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer on (b)(6) 2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
|
| |
|