MAUDE Adverse Event Report: NUVASIVE NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 13600000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product has been returned for investigation as product remains in-situ.No radiographs were provided to confirm the alleged event.A revision procedure is planned but has not occurred.Labeling: ".Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s).".".Post-operative warnings during the postoperative phase it is of particular importance that the physician keeps the patient well informed of all procedures and treatments.Damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications.".".All lock screws should be final-tightened with the counter-torque and torque t-handle.Do not final-tighten through compression instruments (e.G.C/d rack and figure 8 compressor) in the set, as the rod may not be able to normalize to the tulip.Be cautious not to over compress or distract as you can loosen the screws in the spine and potentially pull out the screw.".
 
Event or Problem Description
On (b)(6) 2018, patient underwent a spinal fusion procedure at t9-s2 levels using reline and m5.Post-operative radiographs showed a screw back-out at the right side of s1 level.A revision procedure is planned but has not been performed.
 
Additional Manufacturer Narrative
Supplemental report submitted to update the initial information submitted.
 
Event or Problem Description
Received new information indicating patient underwent a revision procedure on (b)(6) 2018 with no reported events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE RELINE SYSTEM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE
7475 lusk blvd
san diego CA 92121
MDR Report Key7953131
Report Number2031966-2018-00150
Device Sequence Number1190669
Product Code NKB
UDI-Device Identifier00887517551719
UDI-Public887517551719
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K143684
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number13600000
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/18/2018
Supplement Date Received by Manufacturer11/08/2018
Initial Report FDA Received Date10/10/2018
Supplement Report FDA Received Date11/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age58 YR
-
-