• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUBALANCE; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUBALANCE; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUBALANCE
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with no defect found.Returned test strips evaluated with defect found.Qc investigation on (b)(6) 2018 returned test strips produce high readings.Most likely underlying root cause of high results are due to improper storage: vial left open for an extended period of time.(b)(4).Manufacturer contacted customer on (b)(6) 2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 172, 175, 183, 178, 164 mg/dl.Customer's expected fasting blood glucose test result range is 90 - 140 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the bathroom.The test strip lot manufacturer's expiration date is 04/30/2020 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: result 1 :172mg/dl date: (b)(6) 2018 time:08:52am fasting ; result 2 :175mg/dl date: (b)(6) 2018 time:08:15 am fasting ; result 3 :183mg/dl date: (b)(6) 2018 time:08:12am fasting; result 4 :178mg/dl date: (b)(6) 2018 time:10:28am fasting; result 5 :164mg/dl date: (b)(6) 2018 time:08:20am fasting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUBALANCE
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7953280
Report Number1000113657-2018-01018
Device Sequence Number19198730
Product Code NBW
UDI-Device Identifier00021292003135
UDI-Public(01)00021292003135
Combination Product (Y/N)N
Initial Reporter StateAL
PMA/510(K) Number
K090495
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 10/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/30/2020
Device Model NumberTRUBALANCE
Device Catalogue NumberH4H01-81
Device Lot NumberBU5193
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2018
Initial Report FDA Received Date10/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-