| Model Number |
01-SETSCREW |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
| Date of Event |
09/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to rti surgical.A dhr review was conducted and confirmed that this device met manufacturing specifications prior to shipping.The screws all passed thread profile specifications.
|
| |
|
Event or Problem Description
|
|
Patient presented with post operative pain and it was noticed through xray that the screws had loosened and a rod was shifted out of place.A revision surgery was performed on (b)(6) 2018 to replace the rods and screws.
|
| |
|
Search Alerts/Recalls
|