• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125250-28
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined that the reported difficulties and subsequent treatments appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was performed to treat a moderately tortuous and heavily calcified lesion in the mid left anterior descending artery.Following pre-dilatation, a 2.5x28mm xience alpine stent delivery system (sds) was being advanced to the lesion; however, it failed to cross due to the anatomy.An attempt to remove the sds was made, but it was stuck with the heavy calcification and the stent dislodged.A 1.2mm unspecified balloon catheter was advanced and the stent was partially expanded to stabilize it on the guide wire.A snare device was advanced and used to successfully retrieve the stent into the guiding catheter.The dislodged stent and guiding catheter were removed together.The procedure was successfully completed with the deployment of a 3.0x26mm and 3.0x12mm non-abbott stents.There were no adverse patient sequela and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7953432
Report Number2024168-2018-07830
Device Sequence Number19177318
Product Code NIQ
UDI-Device Identifier08717648199196
UDI-Public08717648199196
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 10/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/03/2021
Device Catalogue Number1125250-28
Device Lot Number7121261
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2018
Initial Report FDA Received Date10/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient Weight80
-
-