Model Number ECD02-A
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Laceration(s) of Esophagus (2398)
Date of Event 06/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Notification: a company representative was informed about this incident during a routine in-service visit on (b)(6) 2018 from a nurse in the neurological icu.The nurse said the incident occurred "a few months ago".Device investigation: the device was returned for investigation on (b)(6) 2018 and was inspected by attune medical, senior director of operations, (b)(6).The device showed no signs of damage or deficiency and functioned correctly when connected to a blanketrol ii heat exchanger.The device did not contain any sharp edges, and the stiffness of both the main body of the device and the distal tip were similar to a typical production device.The investigation of the device did not reveal any device defects or non-conformances that may have contributed to the esophageal tear.Clinical investigation: attune medical clinical expert, dr.(b)(6), was consulted on the details of this incident.Because this incident involved a severely injured, intubated, poly-trauma patient with extensive cranio-facial injuries sustained from high-speed blunt trauma, it is difficult to determine both the timing and the etiology of the esophageal injury.The severity of the patient's injuries allow the distinct possibility that esophageal injury, along with other bone or organ injuries seen with severe blunt trauma, occurred as a result of the significant forces sustained during the patient's initial impact.Rapid deceleration is known to result in a variety of internal organ injuries, including solid organs (liver, spleen, kidney, etc.) and hollow organs (intestine, esophagus, bladder, etc.).As such, the patient's esophagus may have been damaged or torn prior to the patient's arrival to the neuro icu.Additionally, the esophageal injury may have been caused by placement of either the endotracheal or the orogastric tube, both of which were placed prior to the ensoetm.Finally, a small esophageal injury may have been exacerbated by placement of the ensoetm into an esophagus with an orogastric tube already in place.Labeling investigation: the dhr for lot #199100918a was reviewed to determine the version of the ifu that was used to manufacture this lot.(b)(4) instructions for use was used to manufacture this lot.(b)(4) contains the following warnings/precautions that may be relevant to this incident: (1) "the ensoetm may cause or exacerbate esophageal tissue damage in patients with known esophageal deformity or evidence of esophageal trauma." and (2) "dual placement of other devices in the esophagus with the ensoetm is an unintended use and may result in patient harm." insertion of the ensoetm into a patient with evidence of esophageal trauma is a known risk, and clinicians are appropriately warned that insertion of the ensoetm into these patients may cause additional harm.Furthermore, dual placement of devices in the esophagus (e.G.Attempting to place an ensoetm in patients with an orogastric tube already in place) is a known risk, and clinicians are warned that this action may cause patient harm.Conclusion the exact cause of, or the timing of, the esophageal injury is unable to be determined from the information available.The esophageal injury may have been present prior to the patient's arrival at the neuro icu, it may have been caused or exacerbated by insertion of the endotracheal tube or orogastric tube, and/or it may have been caused or exacerbated by dual placement of the ensoetm with an orogastric tube.If the ensoetm contributed to the esophageal tear, the cause was most probably user error due to both the insertion of the ensoetm into a patient with severe craniofacial trauma (and the consequent potential for esophageal injury) and the insertion of the ensoetm into a patient with an orogastric tube already in place.
 
Event or Problem Description
The patient arrived at the neuro icu with severe craniofacial trauma described as occipital bone fracture, bilateral sphenoid fracture, first maxillary incisor fracture, temporal bone fracture extending into temporomandibular joint, and nasal bone fractures.These injuries were caused by jumping out of a car traveling approximately 50 mph.The patient was initially intubated with an endotracheal tube and a standard orogastric tube was placed.A nurse subsequently attempted to place the ensoetm without removing the orogastric tube.The nurse experienced some difficulty placing the ensoetm, but was able to successfully place the device.After several hours, the neuro icu staff noticed crepitus in the neck area.Upon inspection, they discovered an injury, described as a "tear" to the esophagus.Medical intervention was required to clip the tear.The patient survived the incident and was later discharged from the neuro icu.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSOETM
Common Device Name
ESOPHAGEAL THERMAL REGULATION DEVICE
Manufacturer (Section D)
ADVANCED COOLING THERAPY INC. D/B/A ATTUNE MEDICAL
3440 s dearborn st #215-south
chicago IL 60616
MDR Report Key7953505
Report Number3010412492-2018-00001
Device Sequence Number1531123
Product Code PLA
UDI-Device Identifier00857915006050
UDI-Public(01)00857915006050(17)200913(10)199100918A
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K172493
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 10/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/13/2020
Device Model NumberECD02-A
Device Lot Number199100918A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2018
Initial Report FDA Received Date10/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age26 YR
-
-