• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SPRINT QUATTRO SECURE; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPRINT QUATTRO SECURE; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number 694758
Medical Device Problem Codes Over-Sensing (1438); Pacing Problem (1439)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/12/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: 5076, implanted: (b)(6) 2016.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the lead integrity alert (lia) was triggered for the right ventricular (rv) lead due to oversensing.During the revision procedure, when the rv lead was disconnected from the device and the rv port was open with the device out of the pocket sitting on the patient, pacing inhibition was noted when the physician touched the can.The rv lead was explanted and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the full lead was returned in segments and analyzed.The analysis indicated that the distal conductor of the lead developed a fracture due to flexing while in vivo.The inner insulation of the lead developed a breach due to abrasion while in vivo.The outer insulation of the lead was breached due to bi/multi-lumen tubing voids while in vivo.The outer insulation was ruptured.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7953525
Report Number2649622-2018-16552
Device Sequence Number1464164
Product Code LWS
UDI-Device Identifier00643169535961
UDI-Public00643169535961
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/28/2018
Device Model Number694758
Device Catalogue Number694758
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2018
Initial Date Received by Manufacturer 09/12/2018
Supplement Date Received by Manufacturer10/11/2018
Initial Report FDA Received Date10/10/2018
Supplement Report FDA Received Date10/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DTBA1Q1, 459888
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age54 YR
Patient Weight118
-
-