Model Number 8225275
Medical Device Problem Codes Use of Device Problem (1670); Device Sensing Problem (2917)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A healthcare professional (hcp) reported via a distributor that after intubation during a thyroidectomy procedure, the probe did not function properly.There was electric current but there was no response.The doctor checked all the connections but was not able to resolve the issue.Another probe was used which functioned well.The procedure was completed but delayed for about 5 minutes.There was no patient impact.
 
Additional Manufacturer Narrative
Analysis found that visually, there was no damage or anomalies in the construction of the device.The tip fit securely in the handle.The resistance of the assembly measurement was 0.2ohms.The probe was plugged into a nim system (with a patient simulator) using 1.0 ma stimulating current and 100uv threshold; when stimulating, the system returned 1.02 ma and a waveform / event tone over 300uv.There was no evidence of improper manufacturing and no malfunction found as it relates to the complaint.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM® PROBE
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7953736
Report Number1045254-2018-00485
Device Sequence Number13503607
Product Code ETN
UDI-Device Identifier00681490176750
UDI-Public00681490176750
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K934426
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/05/2026
Device Model Number8225275
Device Catalogue Number8225275
Device Lot Number0215041108
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2018
Supplement Date Received by Manufacturer10/15/2018
Initial Report FDA Received Date10/10/2018
Supplement Report FDA Received Date11/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-