PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Failure to Sense (1559)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and was unable to duplicate the reported issue.Physio replaced the therapy connector and therapy cable as a precaution.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue couldn't be conclusively determined.
 
Event or Problem Description
The customer contacted physio-control to report that their device was intermittently not reading an ecg signal through the paddles lead and displayed dashed lines.Without this functionality the device would not have the ability to function in aed mode, and the need for defibrillation therapy may not be determined.Physio-control has made multiple attempts to contact the customer in order to obtain additional information about both the patient and the event; however, no response has been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7954022
Report Number0003015876-2018-01682
Device Sequence Number1526144
Product Code MKJ
UDI-Device Identifier00883873884348
UDI-Public00883873884348
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001372
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2018
Initial Report FDA Received Date10/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age45 YR
-
-