| Model Number |
9734410 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The ratcheting handle was returned to the manufacturer for analysis.Analysis found that the end cap had been broken off and was missing.Otherwise, the ratchet and tool retention mechanisms were intact and functional.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the top of the ratcheting handle was missing.There was no patient present when this issue was observed.
|
| |
|