STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL13.2
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lens was returned dry, in a pouch.Visual inspection found one haptic torn.There was evidence of residue on the lens.No similar complaint was reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated the surgeon inserted a 13.2mm micl13.2 implantable collamer lens, -9.5 diopter, and the lens flipped upside down in the patients eye.The surgeon did not like how the lens looked, so decided to remove the lens.The lens was removed within the same surgery with no patient injury.The backup lens was implanted with no problem and the patient is doing well.The reporter was unable to provide any additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7954103
Report Number2023826-2018-01482
Device Sequence Number427815
Product Code MTA
UDI-Device Identifier00841542103282
UDI-Public00841542103282
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Model NumberMICL13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2018
Initial Report FDA Received Date10/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age43 YR
-
-