• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES AORTIC EXCLUDER AAA ENDOPROSTHESIS (LOW PERM); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES AORTIC EXCLUDER AAA ENDOPROSTHESIS (LOW PERM); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number PXT261412
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 10/01/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The review of the manufacturing paperwork verified that this lot met all pre-release specifications.The gore® excluder® aaa endoprosthesis instructions for use (ifu) states that adverse events that may occur and/or require intervention include, but are not limited to endoleak and endoprosthesis component migration.
 
Event or Problem Description
On (b)(6) 2013, the patient underwent an endovascular procedure to repair an abdominal aortic aneurysm using a gore® excluder® aaa endoprosthesis (pxt261412/ (b)(4)).The procedure was completed without any reported issues and the patient tolerated the procedure.In (b)(6) 2017 (date unknown), a follow up examination revealed an endoleak thought to be a proximal type i or type iii endoleak.In (b)(6) 2018 (date unknown), a follow up examination revealed that the endoleak remained.On (b)(6) 2018, an aortic extender component was placed in the proximal part of the trunk-ipsilateral leg component.It was confirmed that the endoleak was resolved.The patient tolerated the procedure.It was reported that the trunk-ipsilateral leg component was thought to be migrated distally (approximately 10 mm) gradually since the proximal neck was angulated.Reportedly the migration contributed to the endoleak.Therefore, the endoleak is thought to be proximal type i instead of type iii.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC EXCLUDER AAA ENDOPROSTHESIS (LOW PERM)
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key7954214
Report Number3007284313-2018-00286
Device Sequence Number427871
Product Code MIH
UDI-Device Identifier00733132610211
UDI-Public00733132610211
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2014
Device Catalogue NumberPXT261412
Device Lot Number9931099
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age83 YR
-
-