MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC S7; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC S7; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9733534
Medical Device Problem Codes No Device Output (1435); Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/14/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation system being used during a functional endoscopic sinus surgery (fess).It was reported that the instrument tracker had recognition failure.Navigation was aborted and there was a surgical delay of less than one hour.There was no reported impact to patient outcome.
 
Additional Manufacturer Narrative
The patient tracker was returned to the manufacturer for analysis.Analysis found that when connected to a known good system the returned tracker was recognized and tracked without issue.No problem found.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S7
Common Device Name
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
louisville
MDR Report Key7954309
Report Number1723170-2018-05069
Device Sequence Number428439
Product Code HAW
UDI-Device Identifier00613994353467
UDI-Public00613994353467
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9733534
Device Catalogue Number9733534
Device Lot Number160407E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2017
Initial Date Received by Manufacturer 11/14/2017
Supplement Date Received by Manufacturer11/14/2017
Initial Report FDA Received Date10/10/2018
Supplement Report FDA Received Date11/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-