| Model Number |
9733534 |
| Medical Device Problem Codes |
No Device Output (1435); Human-Device Interface Problem (2949)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/14/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system being used during a functional endoscopic sinus surgery (fess).It was reported that the instrument tracker had recognition failure.Navigation was aborted and there was a surgical delay of less than one hour.There was no reported impact to patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
The patient tracker was returned to the manufacturer for analysis.Analysis found that when connected to a known good system the returned tracker was recognized and tracked without issue.No problem found.If information is provided in the future, a supplemental report will be issued.
|
| |
|