| Model Number |
MT22608-PNK |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) /2018 that on (b)(6) 2018, the receiver had intermittent audio output.No additional event or patient information is available.No product was provided for evaluation.The complaint confirmation of the receiver having intermittent audio output could not be determined.A probable cause could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The receiver was returned for evaluation.An exterior visual inspection was performed and passed.The receiver log was downloaded and reviewed, finding no errors related to the complaint.A global receiver functional test was performed, and it failed the speaker tests.A global communication tool software test was performed, and it failed sound tests.Manual "try it" testing was performed and failed.The receiver case was opened for an interior inspection and passed.Speaker intermittent testing was performed and failed.Speaker resistance was measured and failed.The reported event of an intermittent audio output was confirmed.The probable cause was determined to be a defective speaker.
|
| |
|