A customer reported a misidentification of klebsiella pneumoniae ssp.Pneumoniae as klebsiella oxytoca when tested with the vitek® 2 gn id card lot# 2410534203 and using vitek 2 software version 8.01.The customer reported the antibiogram matched k.Oxytoca, but testing of a blood culture isolate and re-test of urine isolate from the same patient resulted in an identification of k.Pneumoniae ssp.Pneumoniae.No other reference testing was performed.The customer strain and raw data are not available.The only set up information provided by the customer was that the testing for the urine isolate was from cpse agar and from drigalski for the blood culture isolate.No testing was performed to confirm the identification of k.Pneumoniae ssp.Pneumoniae.Two gn lab reports were submitted.One showed a very good identification of k.Pneumoniae ssp pneumoniae.The other showed an excellent identification of k.Oxytoca with two atypical positive reactions (larl, 5kg) for an identification of k.Pneumoniae ssp.Pneumoniae according to the gn knowledge base.Atypical positive reactions can indicate contamination, mixed culture, use of non recommended media or other user set up errors or an atypical strain.However, without the strain or raw data, it's not possible to further evaluate the cause of the misidentification.The vitek 2 gn lot # 2410534203 met final qc release criteria.This lot passed qc performance testing.
|