| Catalog Number |
254500049 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patella clamp does not grip tight enough when using the clamp and patella drill bit.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary the device associated with this report was not returned for evaluation.The investigation could not verify or identify any product contribution to the reported event with the information provided as the reported device was not returned for evaluation.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate. .
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Upon evaluation of the returned patella/femoral drill, the flutes are worn form normal use.The drill would not have contributed to the reported event.
|
| |
|