| Model Number |
METRIX AIR |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not returned for evaluation.No replacement product needed at this time.Test strip were expired.Customer will purchase new strips.Most likely underlying root cause: mlc-22-client is testing with expired test strips test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the initial concern was resolved- unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for erratic blood glucose test results.Customer stated that it started a week ago.The customer is concerned with tests results from reported back to back blood tests of 330 and 150 mg/dl fasting.The customer's expected fasting blood glucose test result range is 120 - 130 mg/dl.The customer is only concerned with the erratic blood results, not high.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is stored according to specification in the closet.The test strip lot manufacturer's expiration date is 03/28/2018 (test strips are expired at time of call) and open vial date is unknown.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|
Search Alerts/Recalls
|