| Catalog Number |
RBYPOD3 |
| Medical Device Problem Codes |
Device Damaged Prior to Use (2284); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the pod 3 was kinked approximately 5.0 cm from the proximal end of the pusher assembly.Conclusions: evaluation of the returned pod3 revealed a kinked device.If the pod3 is forcefully manipulated or handled during removal from the packaging, damage such as a kink may occur.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure using pod3s.During the procedure, the physician inserted a lantern delivery microcatheter (lantern) into patient¿s body and was checking the lantern positioning on the screen.During the same time, the hospital technologist pulled a pod3 out of its dispenser hoop and noticed the pusher wire was bent; therefore, it was not used in the procedure.It is unclear if the pod3 pusher wire was bent prior to retracting it out of the packaging hoop.The procedure was completed using a new pod3 with the same lantern.
|
| |
|