Catalog Number 03.617.931
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional narrative: initial reporter is synthes sales consultant.A review of the device history records has been requested.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2018, during zero-profile variable angle(va) procedure, the handle of angled awl came off behind the case after the surgeon used the mallet on the back.During the procedure, quick connect of hex adaptor broke as it was entered in the screw.The quick connect shaft that goes into the ratcheting screwdriver handle broke were it connect so that it cannot enter in the handle.The broken piece is still inside the handle and cannot be removed.There was no surgical delay reported.Procedure was successfully completed.It came out of patient fine yet fell off when passing it off.Patient was not harmed due to malfunction.Concomitant device reported: screw (part unknown, lot unknown, quantity 1), mallet (part unknown, lot unknown, quantity 1).This report is for one (1) hex adaptor.This is report 2 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Device history lot : part: 03.617.931, lot: 3474355, manufacturing site: (b)(4), release to warehouse date: 27.July 2010.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacturing process.Only top level of the device history record reviewed as sub-components are not lot tracked.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Investigation summary.The broken hex tip of the initially returned hex quick connect adapter was returned lodged with ratcheting screw driver handle used with it.Therefore the visual inspection and investigation details have been updated as below: device condition: the returned devices showed broken shaft portion on the coupling shaft that assembles with the screwdriver handle.The breakage was observed on the groove portion of the shaft which is the smallest cross section of the shaft.The broken shaft portion was also returned.No new issues were identified.With the damage identified the returned device, the device condition is consistent with the given complaint condition and therefore the complaint condition was confirmed.Dhr review performed showed no issues relevant to complaint condition.Material review was not possible as sub-components are not lot tracked.However, upon visual inspection on both the broken end regions, no material inconsistencies were observed.With the breakage observed on the smallest cross-section it is possible that the device would have been subjected to unintended excessive usage forces when used with handle.No definitive root cause could be determined from the available information.However, during the investigation no design/manufacturing issues were identified.Therefore, based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Concomitant device reported: screw (part unknown, lot unknown, quantity 1), mallet (part unknown, lot unknown, quantity 1), ratcheting screwdriver handle (part 311.023, lot t997332, quantity 1).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A device history record (dhr) review was conducted: part: 03.617.931 , lot: 3474355 , manufacturing site: hägendorf , release to warehouse date: 27.July 2010.The device history record shows this lot of (b)(4) pieces was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and functional criteria at the time of release with no issues documented during the manufacturing process.Only top level of the device history record reviewed as sub-components are not lot tracked.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.A product investigation was conducted.Customer quality investigation: device condition: the returned devices showed broken shaft portion on the coupling shaft that assembles with the screwdriver handle.The breakage was observed on the groove portion of the shaft which is the smallest cross section of the shaft.The broken shaft portion was not returned.No new issues were identified.With the damage identified the returned device is consistent with the given complaint condition and therefore the complaint condition was confirmed.Drawing/document specification review: relevant product drawings were reviewed during the investigation and no design issues were noted.Dimensional analysis was not performed on the relevant features identified with post manufacture damage.Dhr review performed showed no issues relevant to complaint condition.Material review was not possible as sub-components are not lot tracked.However, upon visual inspection on the broken region no material inconsistencies were observed.With the breakage observed on the smallest cross-section it is possible that the device would have been subjected to unintended excessive usage forces when used with handle.No definitive root cause could be determined from the available information.However, during the investigation no design/manufacturing issues were identified.Therefore, based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEX ADAPTOR-QC
Common Device Name
MISC ORTHO SURGICAL INSTR
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7958001
Report Number2939274-2018-54344
Device Sequence Number1301624
Product Code LXH
UDI-Device Identifier10705034717272
UDI-Public(01)10705034717272
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.617.931
Device Lot Number3474355
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2018
Initial Date Received by Manufacturer 09/25/2018
Supplement Date Received by Manufacturer10/12/2018
11/15/2018
12/05/2018
Initial Report FDA Received Date10/11/2018
Supplement Report FDA Received Date10/29/2018
11/16/2018
12/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-