• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESHAPE LIFE SCIENCES MAESTRO RECHARGEABLE SYSTEM; RECHARGEABLE NEUROREGULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESHAPE LIFE SCIENCES MAESTRO RECHARGEABLE SYSTEM; RECHARGEABLE NEUROREGULATOR Back to Search Results
Model Number 2002
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Codes Failure of Implant (1924); Therapeutic Response, Decreased (2271)
Date of Event 06/04/2018
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2018, patient 835-bam at skyridge experienced a red flashing light on the mc when connecting with the rnr at a post-op visit.The reshape field specialist tried several times to clear the flashing red lights when connected to the cp, however, the cp "locked up." the field specialist then hooked up another rnr (implant kit) to the cp, the red light disappeared and a green flashing light appeared on the mc signaling the patient was receiving therapy.The rnr logs were unable to upload due to the lock out on the cp, however, debug and data files were sent for investigation.On (b)(6) 2018, the field specialist met with the patient to attempt to connect the device to the cp again to clear the alarms.After several attempts, the cp was remained locked.The field specialist expressed to the implanting physician if they are unable to resolve this issue, revision surgery may be necessary.On (b)(6) 2018, the implanting physician performed revision surgery, removed the rnr and replaced with a new rnr.The device performed as expected and patient is doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAESTRO RECHARGEABLE SYSTEM
Common Device Name
RECHARGEABLE NEUROREGULATOR
Manufacturer (Section D)
RESHAPE LIFE SCIENCES
2800 patton road
saint paul MN 55113
MDR Report Key7958255
Report Number3005025697-2018-00010
Device Sequence Number1386604
Product Code PIM
UDI-Device Identifier00857334004262
UDI-Public00857334004262
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P130019
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 10/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number2002
Device Catalogue Number2002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2018
Initial Report FDA Received Date10/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-