| Catalog Number |
170505 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
|
| |
|
Event or Problem Description
|
|
Screw hole stripped on femoral trial.Case type: pka.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding crack/fracture involving a mako trial was reported.The event was confirmed.Method & results: device evaluation and results: visual inspection and material analysis: an inspection of the returned device by a materials analyst noted "damage observed on threads consistent with crossthreading and off axis loading.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined." medical records received and evaluation: n records were made available for evaluation.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there have been no other similar reported events for the reported lot.Conclusions: no further investigation is required at this time.The root cause determined because insufficient information was provided.Further information such as operative notes are needed to investigate this event further.If additional information becomes available, this investigation will be reopened.
|
| |
|
Event or Problem Description
|
|
Screw hole stripped on femoral trial.Case type: pka.
|
| |
|