| Catalog Number |
398.754 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Misassembly by Users (3133)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.A review of the device history records has been requested.Device was used for treatment, not diagnosis.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2018, during an unknown procedure, the pelvic arm could only be removed by force after connecting the collinear reduction clamp sliding mechanism.It was unknown if there was a procedure or patient involvement.Concomitant devices reported: percutaneous arm, 255mm (part #: 398.754, lot #: 1515783, quantity: 1), bone hook-sape arm, 206mm (part #: 398.756, lot #: 10-9537, quantity: 1) and hohmann-style arm, 183mm (part #: 398.752, lot #: 1515781, quantity: 1).This complaint involves four (4) devices.This report is 2 of 4 for (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6), 2018, the pelvic arm, percutaneous arm, bone hook-shape arm and hohmann-style arm were difficult to remove after connecting from the collinear reduction clamp sliding mechanism.The pelvic arm was removed by force.There was no patient involvement.This report is for a percutaneous arm 255mm.There are no concomitant devices.
|
| |
|
Additional Manufacturer Narrative
|
|
Reporter is a synthes employee.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Part: 398.754; lot: 1515783; manufacturing site: selzach; supplier: leitner ag; release to warehouse date: november 14, 2016.The device history record shows this lot was processed through the normal.Manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and functional criteria at the time of release with no issues documented during the manufacturing process.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|