• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX ENDOTRACHEAL TUBE; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX ENDOTRACHEAL TUBE; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 100/133/075
Medical Device Problem Codes Deflation Problem (1149); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Three portex® endotracheal tubes were returned for analysis in used conditions.No discrepancies were found upon visual inspection.Functional testing was performed by inflating the syringe while submerged under water; leakage was observed coming from a small tear in the cuff to all 3 samples.Relevant documents were reviewed and deemed adequate.An audit of the production floor was performed by reviewing training records and relevant procedures; all deemed adequate.Thirty two samples were reviewed for tracheal inflation line and leak tests; no discrepancies or leaks were found.Based on the evidence the complaint was confirmed.However, the root cause is unknown.
 
Event or Problem Description
Information was received indicating that immediately following intubation of the patient with a smiths medical portex endotracheal tube, air came out of the cuff.There were no reported adverse patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ENDOTRACHEAL TUBE
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim
6000 nathan ln n
minneapolis, MN 55442
7633833310
MDR Report Key7958675
Report Number3012307300-2018-04244
Device Sequence Number5522665
Product Code BTR
UDI-Device Identifier15019315020360
UDI-Public15019315020360
Combination Product (Y/N)Y
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date08/02/2022
Device Model Number100/133/075
Device Catalogue Number100/133/075
Device Lot Number3463144
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2018
Supplement Date Received by Manufacturer12/03/2018
12/03/2018
Initial Report FDA Received Date10/11/2018
Supplement Report FDA Received Date10/11/2018
01/03/2019
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/31/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-