| Model Number |
100/133/075 |
| Medical Device Problem Codes |
Deflation Problem (1149); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Three portex® endotracheal tubes were returned for analysis in used conditions.No discrepancies were found upon visual inspection.Functional testing was performed by inflating the syringe while submerged under water; leakage was observed coming from a small tear in the cuff to all 3 samples.Relevant documents were reviewed and deemed adequate.An audit of the production floor was performed by reviewing training records and relevant procedures; all deemed adequate.Thirty two samples were reviewed for tracheal inflation line and leak tests; no discrepancies or leaks were found.Based on the evidence the complaint was confirmed.However, the root cause is unknown.
|
| |
|
Event or Problem Description
|
|
Information was received indicating that immediately following intubation of the patient with a smiths medical portex endotracheal tube, air came out of the cuff.There were no reported adverse patient effects.
|
| |
|