• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC DAVINCI XI; SURGEON SIDE CONSOLE, SMART PEDALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC DAVINCI XI; SURGEON SIDE CONSOLE, SMART PEDALS Back to Search Results
Model Number 380677-10
Device Problem Poor Quality Image (1408)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/18/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4) - intuitive surgical, inc.(isi) received both the hrsv and the pmsc involved with this complaint.Failure analysis was able to confirm the reported failure on the pmsc.The pmsc was installed into a test system and there was video from the ssc and the vision side console.However, the video output could not be tested due to the dvi connector having bent pins.The failure analysis for the hsrv has not been completed and is still in progress.A follow-up mdr will be submitted once the device evaluation has been completed or if additional information is received.The complaint is being reported due to a da vinci system malfunction rendering the da vinci system unavailable for use after the start of a surgical procedure.Although no patient harm occurred, if the reported malfunction were to recur, it could cause or contribute to an adverse event.
 
Event Description
It was reported that during a da vinci-assisted sigmoid colectomy procedure, no vision was observed in the left eye.The customer tried a new scope but the issue was not resolved.The intuitive surgical, inc.(isi) technical service engineer (tse) advised to power cycle the system and cycle the breaker on the back of the surgeon side console (ssc) and remove the cable from the back of the ssc but the issue persisted.At that time, the surgeon made the decision to convert and complete the procedure using traditional laparoscopic techniques.There was no report of patient harm, adverse outcome or injury.An isi field service engineer (fse) was dispatched to the facility and was able to reproduce the reported failure.To resolve the issue, the fse replaced the left high resolution stereo viewer (hsrv) and the personality module surgeon console (pmsc).The hrsv is located on the surgeon side console and provides the 3d image captured from the endoscope inserted in the patient.The pmsc is a module that consists of 5 interface boards.
 
Manufacturer Narrative
Intuitive surgical, inc.(isi) has completed the device evaluation for the high resolution stereo viewer.Failure analysis confirmed the customer reported failure mode.There was no image due to the main board having a defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAVINCI XI
Type of Device
SURGEON SIDE CONSOLE, SMART PEDALS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
950 kifer rd.
sunnyvale CA 94086
MDR Report Key7959062
MDR Text Key123521377
Report Number2955842-2018-10601
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number380677-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2018
Date Manufacturer Received12/05/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES; DA VINCI INSTRUMENTS AND ACCESSORIES
-
-