| Model Number |
ZCT150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Bag Tear (2639)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: unknown, information not provided.Gender/sex: unknown, information not provided.Date of event: unknown, not provided.Implant and explant dates: if implanted or explanted; give date: not applicable as the lens was inserted and removed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the tecnis toric intraocular lens (model zct150, +15.5 diopter) was implanted, removed and replaced in the same procedure as the patient moved and the capsule ruptured.Reportedly, a three-piece intraocular lens was implanted as a replacement.No incision enlargement, no vitrectomy was performed and no patient injury was reported.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation: yes.Returned to manufacturer on: 11/14/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.A visual inspection of the return shows the product was cut in half, most probably to make removal possible.Considering the condition of the lens, further analysis is not possible.The complaint event could not be confirmed.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order number was performed.The search revealed that there was one other complaint received for this production order number however there is no relation between these complaints.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.A search revealed no other related complaints were received under the production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|