• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. OES CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/11/2018
Event Type  malfunction  
Event Description
Olympus medical systems corp.(omsc) was informed that as a result of multiple microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device: [first time; (b)(6) 2018] unspecified microbes (2cfu / 100ml), [second time; (b)(6) 2018] unspecified microbes (2cfu / 100ml).The device had been manually reprocessed using peracetic acid.There was no report of infection associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device has not been returned to olympus medical systems corp.(omsc) but was returned to olympus (b)(4).(b)(4) sent the subject device to a third party laboratory for microbiological testing.As a result of the testing, the sample collected from all channels of the subject device tested positive for gram-positive bacteria (1cfu/endoscope).The testing result cleared the (b)(4) guideline.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
MDR Report Key7960028
MDR Text Key126079693
Report Number8010047-2018-01977
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/11/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-