| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a power (battery life) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long-term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 17-oct-2018 with the following findings: a review of the pump black box data showed no evidence of short battery life; the user was installing discharged batteries.During a visual inspection of the pump, the battery compartment was observed to be intact with no damage or evidence of moisture ingress.The battery cap secured properly to the pump and a power loss was not observed.Current draws measured within specifications.The complaint of a battery life issue was not duplicated during investigation.
|
| |
|