| Catalog Number |
05122287001 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event occurred in (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer complained of discrepant sodium (na) results for 1 patient tested on a cobas b 123 system compared to a beckman coulter analyzer.The na result from the beckman coulter analyzer at 1:59 p.M.Was 162 mmol/l.A sample was obtained at 1:59 p.M.For the cobas b 123 instrument and the na result at 2:09 p.M.Was 152 mmol/l.The customer performed comparison testing with the patient sample for troubleshooting purposes on 4 other cobas b123 instruments: cobas b 123 #1 at 2:11 p.M.Was 156 mmol/l, cobas b 123 #2 at 2:17 p.M.Was 159 mmol/l, cobas b 123 #3 at 2:31 p.M.Was 161 mmol/l, cobas b 123 #4 at 2:37 p.M.Was 161 mmol/l.The customer is concerned with the lower na result of 152 mmol/l from the b 123 instrument in question.There was no allegation that an adverse event occurred.Qc and calibration were acceptable.The na electrode lot number and expiration date were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The log files were investigated and it was determined the customer did not apply the recommended correlation factors when comparing measurements from cobas b 123 to a lab analyzer.Retention testing of the lot of na sensors involved met specifications.The investigation did not identify a product problem.A specific root cause could not be determined.
|
| |
|