Catalog Number 623-10-32E
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history records indicates that the reported devices were manufactured and accepted into final stock with no reported discrepancies.The complaint history review indicated that there were no similar events for the reported lot.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Customer reported that during a thr, the surgeon stephen had implanted a tritanium primary shell, however when he tried to impact the 32 10 degree trident liner into the shell, it wouldn't lock in.Another implant from the shelf was used to complete the case and there was only a delay of few minutes.
 
Additional Manufacturer Narrative
An event regarding seating/locking issue involving a trident liner was reported.The event was not confirmed.Method & results: product evaluation and results: examination of the returned device by an engineer noted the following: damage observed on outer rim of insert consistent with implantation and explantation process.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Medical records received and evaluation: not performed as no medical information was received.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there were no other events for the lot provided.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time as insufficient information was received by stryker orthopaedics.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Customer reported that during a thr, the surgeon stephen had implanted a tritanium primary shell, however when he tried to impact the 32 10 degree trident liner into the shell, it wouldn¿t lock in.Another implant from the shelf was used to complete the case and there was only a delay of few minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 10° X3 INSERT 32MM ID
Common Device Name
HIP JOINT METAL/POLYMER/METAL SEMI-CONSTRAINED POROUS-COATED UNCEMENTED PROSTHES
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7960716
Report Number0002249697-2018-03310
Device Sequence Number1572423
Product Code LPH
UDI-Device Identifier07613327039856
UDI-Public07613327039856
Combination Product (Y/N)N
Initial Reporter CountryGB
PMA/510(K) Number
K033716
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Catalogue Number623-10-32E
Device Lot NumberJ47AT0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2018
Initial Date Received by Manufacturer 09/14/2018
Supplement Date Received by Manufacturer02/04/2019
Initial Report FDA Received Date10/12/2018
Supplement Report FDA Received Date03/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-