| Model Number |
M00583190 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a truetome dreamwire 44 was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2018.According to the complainant, during the procedure, when the device was cannulated in order to enter the bilde duct, the cutting wire broke.Reportedly, no part of the device detached inside the patient.The procedure was completed with a second truetome dreamwire 44, using the same generator and active cord.There were no user injuries nor patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a truetome dreamwire 44 was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2018.According to the complainant, during the procedure, when the device was cannulated in order to enter the bilde duct, the cutting wire broke.Reportedly, no part of the device detached inside the patient.The procedure was completed with a second truetome dreamwire 44, using the same generator and active cord.There were no user injuries nor patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Visual examination of the returned device revealed that the cutting wire was broken and blackened.There was no other issue noted.The complaint was consistent with the reported event of cutting wire broke.It is most likely that a peak of voltage could have caused the failures noted.Therefore, the most probable root cause is cause traced to component failure since the event was due to expected or random component failure without any design or manufacturing issue.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release to distribution.
|
| |
|