• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN SCRW DIS FEM WDG SZ4 5MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN SCRW DIS FEM WDG SZ4 5MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 71421661
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Injury (2348)
Date of Event 09/05/2014
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a left knee arthroscopy was performed.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.The clinical/medical team concluded, arthrofibrosis is a condition that sometimes occurs in a knee joint that has recently been injured.Risk factors for arthrofibrosis include replacement knee surgery and arthroscopic surgery.Based on the information provided and without the return of the explant for analysis the root cause of the loosening component and arthrofibrosis cannot be concluded.However, the patient falls prior to the revision could not be ruled out as contributing factors to the reported event.The patient impact beyond the reported revision cannot be determined.No further clinical assessment is warranted at this time.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed parts revealed no prior complaints for the listed batches.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN SCRW DIS FEM WDG SZ4 5MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7961977
Report Number1020279-2018-02181
Device Sequence Number16564433
Product Code JWH
UDI-Device Identifier03596010544421
UDI-Public03596010544421
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K041106
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/07/2022
Device Catalogue Number71421661
Device Lot Number11JSM0402B
Initial Date Received by Manufacturer 09/14/2018
Supplement Date Received by Manufacturer09/14/2018
Initial Report FDA Received Date10/12/2018
Supplement Report FDA Received Date05/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age65 YR
-
-