| Catalog Number |
03P93-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Codes |
Congestive Heart Failure (1783); Renal Failure (2041); Respiratory Distress (2045); Chronic Obstructive Pulmonary Disease (COPD) (2237)
|
| Date of Event |
10/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, abbott point of care (apoc) was contacted by a customer regarding i-stat bnp cartridges that yielded a suspected discrepant troponin result on a (b)(6) female patient with chf, copd, chronic kidney disease.There was no additional patient information at the time of this report.Return product is not available.Method: i-stat, date: (b)(6) 2018, result: 1116; lab, (b)(6) 2018, 406.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident #: (b)(4).The investigation was completed on 11/08/2018.A review of the device history record (dhr) confirmed the cartridge lot met finished goods (fg) release criteria.Retained cartridge testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.Ac (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified for bnp cartridge lot d18143a.
|
| |
|