ABBOTT POINT OF CARE I-STAT BNP CARTRIDGE Back to Search Results
Catalog Number 03P93-25
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Codes Congestive Heart Failure (1783); Renal Failure (2041); Respiratory Distress (2045); Chronic Obstructive Pulmonary Disease (COPD) (2237)
Date of Event 10/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2018, abbott point of care (apoc) was contacted by a customer regarding i-stat bnp cartridges that yielded a suspected discrepant troponin result on a (b)(6) female patient with chf, copd, chronic kidney disease.There was no additional patient information at the time of this report.Return product is not available.Method: i-stat, date: (b)(6) 2018, result: 1116; lab, (b)(6) 2018, 406.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident #: (b)(4).The investigation was completed on 11/08/2018.A review of the device history record (dhr) confirmed the cartridge lot met finished goods (fg) release criteria.Retained cartridge testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.Ac (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified for bnp cartridge lot d18143a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT BNP CARTRIDGE
Common Device Name
BNP CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key7962086
Report Number2245578-2018-00260
Device Sequence Number13494115
Product Code NBC
UDI-Device Identifier10054749000217
UDI-Public10054749000217
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K053597
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/13/2018
Device Catalogue Number03P93-25
Device Lot NumberD18143A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/09/2018
Supplement Date Received by Manufacturer11/08/2018
Initial Report FDA Received Date10/12/2018
Supplement Report FDA Received Date11/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age79 YR
-
-