| Model Number |
TRUEMETRIX |
| Medical Device Problem Codes |
Unable to Obtain Readings (1516); High Test Results (2457)
|
| Health Effect - Clinical Codes |
Fatigue (1849); Urinary Frequency / Polyuria (2275); Polydipsia (2604)
|
| Date of Event |
09/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not returned for evaluation.No replacement product needed at this time.Product is working as designed.Most likely underlying root cause: mlc-18- user has high glucose value.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Unable to send product notification letter as no address on file for customer.
|
| |
|
Event or Problem Description
|
|
Customer called in requesting to know what e-5 meant.The e-5 error occurred when the blood sample was absorbed.Informed customer that it is due to a test strip error or that his blood sugar could be above 600mg/dl.As soon as we informed the customer it could be above 600mg/dl customer states, that's it.Asked customer if he had any symptoms of high blood sugar as i listed them customer states he has frequent urination, excessive thirst, and feels tired.Instructed customer to call 911 and seek medical attention.Informed customer we would call him back tomorrow.
|
| |
|