This report was originally filed under medwatch 9611594-2018-00112 on 16 aug 2018.However, it was determined that the manufacturer number was used in error, and the correct manufacturer reference number is (b)(4).This new initial report 3011270181-2018-00043 replaces the original initial report 9611594-2018-00112 and all further updates will be filed under this number.One used sample was received for evaluation.The sample was visually inspected and no obvious damage or defect was noted.The cuff was inflated with air and submerged in water.No leakage was seen.The cuff, pilot balloon, and inflation line were all squeezed and manipulated, and no leakage was observed at any point along the inflation pathway.The cuff maintained inflation throughout the evaluation.The complaint could not be confirmed as reported.No root cause could be determined.All information reasonably known as of 27 sep 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
|