U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. NIM-ECLIPSE® PREAMPLIFIER; STIMULATOR, ELECTRICAL, EVOKED RESPONSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. NIM-ECLIPSE® PREAMPLIFIER; STIMULATOR, ELECTRICAL, EVOKED RESPONSE Back to Search Results
Model Number 945DAQ916
Medical Device Problem Codes Use of Device Problem (1670); Impedance Problem (2950)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis found that the device came in with excessive blood around the seals which leaked onto the top board microprocessor which caused the failure.There were components failure and microprocessor liquid damage.The device was cleaned, replaced top board, applied rtv to connectors, applied tape, replaced 2 torn gaskets, placed into burn-in for 8 hours to check for any heat related failures and tested in accordance with manufacturing specifications.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care provider (hcp) reported that the device failed the impedance test and was not working properly.There was no patient impact reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM-ECLIPSE® PREAMPLIFIER
Common Device Name
STIMULATOR, ELECTRICAL, EVOKED RESPONSE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7962826
Report Number1045254-2018-00504
Device Sequence Number11197408
Product Code GWF
UDI-Device Identifier00643169071933
UDI-Public00643169071933
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K050798
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number945DAQ916
Device Catalogue Number945DAQ916
Device Lot Number205892585
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2018
Initial Report FDA Received Date10/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-