• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON AND JOHNSON SURGICAL VISION TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON AND JOHNSON SURGICAL VISION TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age/date of birth: unknown, information not provided.Gender/sex: unknown, information not provided.Implant and explant dates: if implanted or explanted, give date: not applicable; there was no indication the lens was implanted; therefore, not explanted.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that intraocular lens had a bent haptic.Reportedly, the lens had patient contact.No further information provided.
 
Additional Manufacturer Narrative
Device evaluation: the returned pcb00 device was received in the packaging box.The plunger was in completely advanced position and locked.The pcb00 device was observed under microscope and a scarce amount of lubricant residue was observed inside the device.The lens was observed stuck and out at the end of the cartridge tip with the lead haptic out of the tip with the pushrod adhered.The complaint was not verified in the sample return; however, a lead haptic not folded condition was observed and could be related to the scarce amount of viscoelastic.No product quality deficiency was identified.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION
santa ana CA
MDR Report Key7963249
Report Number2648035-2018-01321
Device Sequence Number202000
Product Code HQL
UDI-Device Identifier05050474558168
UDI-Public(01)05050474558168(17)200116
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/16/2020
Device Model NumberPCB00
Device Catalogue NumberPCB0000150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 09/14/2018
Supplement Date Received by Manufacturer10/23/2018
10/25/2020
Initial Report FDA Received Date10/12/2018
Supplement Report FDA Received Date11/13/2018
11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-