Catalog Number 4002C2060
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The embolization coil was intact with its pusher assembly.The pc400 was advanced out of its introducer sheath with minimal resistance.Then the pc400 was advanced through a demonstration microcatheter and out of the distal tip with minimal resistance.Conclusions: evaluation of the returned pc400 revealed that the device was functional.During functional testing, the pc400 was advanced out of its introducer sheath with minimal resistance.Then the pc400 was advanced through a demonstration microcatheter and out of the distal tip with minimal resistance.The lantern identified in the complaint was not returned to penumbra for evaluation.Therefore, the root cause of the reported resistance and advancement issues could not be determined.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra coil 400s (pc400s).During the procedure, the physician successfully placed four pc400s using a lantern delivery microcatheter (lantern).The physician then felt resistance and was unable to advance another pc400 more than 10 cm into the lantern.The pc400 was therefore removed, and the procedure was completed using another pc400 and the same lantern.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7963378
Report Number3005168196-2018-02031
Device Sequence Number19267382
Product Code HCG
UDI-Device Identifier00814548010762
UDI-Public00814548010762
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/14/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/18/2026
Device Catalogue Number4002C2060
Device Lot NumberF84163
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2018
Initial Report FDA Received Date10/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-